Alternative title | Calibrations for resistance welding equipment with high pulsed currents |
---|
Download | - View accepted manuscript: Calibrations for resistance welding at high pulsed currents (PDF, 1.3 MiB)
|
---|
DOI | Resolve DOI: https://doi.org/10.1109/CPEM.2014.6898234 |
---|
Author | Search for: Djokic, Branislav1 |
---|
Affiliation | - National Research Council of Canada. Measurement Science and Standards
|
---|
Format | Text, Article |
---|
Conference | 29th Conference on Precision Electromagnetic Measurements, CPEM 2014, 24 August 2014 through 29 August 2014 |
---|
Subject | AC current measurement; Digital sampling; Measurement uncertainty; Pulsed currents; Rogowski Coil; Weld current; Calibration |
---|
Abstract | Rogowski coils are widely used as high pulsed current sensors in resistance welding. Their operating conditions reflect on requirements for their calibration. A new calibration system has been developed at NRC Canada for calibrating current sensing coils and weld current monitors for AC resistance welding. The calibration system, which includes a high current source and a coaxial copper cage, has been evaluated at peak pulsed currents of up to 80 kA, at 60 Hz. The best expanded measurement uncertainties (k=2) have been estimated to be within 250 μA/A for magnitude and 175 μrad for phase. |
---|
Publication date | 2014-08 |
---|
In | |
---|
Series | |
---|
Language | English |
---|
Peer reviewed | Yes |
---|
NPARC number | 21272955 |
---|
Export citation | Export as RIS |
---|
Report a correction | Report a correction (opens in a new tab) |
---|
Record identifier | 36a1bf98-80c6-4b32-a05f-31754a18cd1d |
---|
Record created | 2014-12-03 |
---|
Record modified | 2020-06-04 |
---|